BiPAP® S/T System with Encore® Pro SmartCard®

Specifications


Patient Data Range
OUTPUT 4 to 30 cm H2O
IPAP 4 to 30 cm H2O* ± 5 cm H2O **
EPAP 4 to 25 cm H2O* ± 5 cm H2O **
CPAP 4 to 20 cm H2O ± 5 cm H2O **
BREATH RATE 0 to 30 BPM
Greater of ± 1 BPM or ± 10% of the setting (when measured over a 4 minute period)
OPERATING TEMPERATURE 41°F to 95 °F
STORAGE TEMPERATURE -4°F to 140°F
ATMOSPHERIC PRESSURE 5600 feet to sea level
RELATIVE HUMIDITY 15 to 95% (non-condensing)
Ramp Duration 0 to 45 minutes ± 10% of the setting
Dimensions
Size: 9.75" L x 6.625" W x 4.4" H
Weight: 4 lbs approximate


* Limited to 20 cm H2O when using the Bi-Flex feature in S mode.

** Dynamic pressure accuracy is ± 5 cm H2O measured at the patient end of the circuit with a Whisper Swivel II and varying flow conditions.